Page 23 of 63 Pages
16. Liquids
16.1 When sterilizing liquids, it is essential to leave the multi-purpose
knob on sterilization position after the sterilization cycle is over,
letting the steam escape slowly.
16.2 It is recommended to put the bottles on an un-perforated tray,
which can be ordered at Tuttnauer, to prevent the clogging of the
vessel opening in case the liquid starts boiling.
16.3 When the pressure decreases to “0” open the door and unload the
autoclave.
16.4 Turn the multi-purpose valve knob to 0.
Warning
Multi-purpose valve knob should be turned in a clock-wise
direction only.
Caution
If the door handle cannot be turned counter-clockwise for the door
to be opened, the reason may be that the pin is set firmly in the
groove. In order to release the pin, slightly turn the handle in the
closing direction (clockwise). The pin is released, enabling the
operator to open the door. In case the door still cannot be opened,
refer to the instructions in the chapter “Pressure Door Lock
System” (para. 11.11).
NOTE:
After operating the sterilizer, brown stains might appear on the bottom of
the chamber. These stains are a result of the heating elements that are
located at the lower external part of the chamber. The brown color is a
common phenomenon, can be easily be removed, and will not have any
effect on the steriliszed goods.
At the end of the working day turn the main switch to "STOP" position.