Heat T.O. Maximum time for heating the chamber for starting sterilization
When the duration of the heating stage is longer than Heat T.O., the program will stop
and display the message ‗Low Heat‘.
Entry code
11
Resolution
1 minute
Minimum value
20 minutes
Maximum value
60 minutes
Ex Mode The method for exhausting the steam at the end of the process
Entry code
11
Ex Mode
Fast Ex
1
N.A.
2
Slow Ex. (waste)
3
Slow Ex. (liquids)
4
Cooling with compressed air
5
Cooling with water circulation
6
EndTemp Ending Temperature (Exhaust Stage)
Defines the temperature at which the process ends for opening the door.
Entry code
11
Resolution
Minimum value
40ºC (104ºF)
Maximum value
137ºC (278ºF)
PulsNum. No. of pulses in the pre-vacuum stage
Entry code
11
Resolution
1
Minimum value
1
Maximum value
10
VacDip1 Vacuum value in the first pulse
Defines the vacuum value in pulse No.1 of the pre-vacuum stage
Entry code
11
Resolution
1 kPa (0.1 psi)
Minimum value
1 kPa (0.15 psia)
Maximum value
75 kPa (10.8 psia)