OPERATION
&
MAINTENANCE
MANUAL
Pre-Vacuum Steam Heated Autoclave
with one Vertical Sliding Door
and a 36 kW Steam Generator
Model 6671130 2VEP
S.N: 1102058
Cat. No.: MAN205-0195042EN Rev A Aug. 2011
Tuttnauer. Ltd., Har Tuv Industrial Zone, B P.O. Box 170, Beit Shemesh, 99000, Israel, Tel:972 2 9904611
Fax:972 2 9904730
Tuttnauer Europe b.v., Hoeksteen 11, 4815 PR Breda Breda, P.O. Box 7191, 4800 GD Breda, Netherlands.
+31/76-5423510, Fax: +31/76-5423540

Table of Contents

Questions and Answers

  • M
    Michael TannerSep 13, 2025
    Why is the air inlet not working on my Tuttnauer Steriliser?
    • B
      Brian MyersSep 14, 2025
      If the air inlet on your Tuttnauer Steriliser isn't working, it's possible that the air inlet valve is clogged or damaged, requiring repair or replacement. Alternatively, the air filter might be clogged and needs replacement.
  • G
    gabriel39Sep 15, 2025
    What to do if the Tuttnauer Steriliser control unit won't initialize?
    • D
      Dr. Andrew Hernandez MDSep 16, 2025
      If the control unit of your Tuttnauer Steriliser isn't initializing, and there's no display or lights on the front panel, it could be due to the control circuit breaker being switched off (check and switch it on). Alternatively, the power switches on the front panels of the autoclave or on the steam generator might be defective (replace the defective switch).
  • B
    Brenda MillerSep 18, 2025
    What causes the air detector to fail on a Tuttnauer Steriliser?
    • C
      Carly WillisSep 18, 2025
      If the air detector fails on your Tuttnauer Steriliser, it could be due to improper steam quality (check the steam quality and non-condensable gases). Alternatively, the induced air valve might be open or leaking (close or repair the induced air valve).
  • C
    cassandra08Nov 19, 2025
    Why is the steam generator not working on my Tuttnauer Steriliser?
    • G
      gentrycoreyNov 19, 2025
      If the steam generator on your Tuttnauer Steriliser isn't working, several issues could be the cause. The water pump might be defective or blocked, requiring repair or replacement. There could be a blockage or low water pressure in the pump circuit, so you should adjust the water pressure and ensure free water circulation. A tripped circuit breaker due to overload or damage can also be a factor; check for and remove any overload, or replace the circuit. Other potential causes include a short-circuited upper electrode (clean or replace it), a faulty electronic pressure switch or measuring circuits (check, calibrate, or replace them), a dirty or rusted lower electrode (clean or replace it), tripped or damaged circuit breakers (turn off or replace them), a damaged contactor controlling the he...
  • J
    Janice MerrittNov 21, 2025
    What to do if the pre-vacuum phase is not working on my Tuttnauer Steriliser?
    • J
      Jennifer ColemanNov 21, 2025
      If the conditioning phase (pre-vacuum) isn't working on your Tuttnauer Steriliser, and there's no steam inlet or low steam pressure, here's what to check: * Determine if the issue stems from the pump or its electrical supply circuit. * The vacuum valve passing water through the vacuum pump might be stuck or damaged; inspect, repair, or replace it. * The exhaust to pump valve could be stuck or damaged; inspect, repair, or replace the fast exhaust valve. * Ensure the door is properly closed, and the gasket isn't damaged or incorrectly seated; close the door, and fix or replace the gasket if needed. * Check for leakage at the vacuum or exhaust piping; tighten fitting joints and replace defective packing. * The steam inlet valve might be blocked or damaged; fix or replace the stea...