9.3 Replacing the Safety Valve
Caution!
Before starting, be sure that the electric cord is disconnected and that
there is no pressure in the autoclave.
Note:
These instructions are valid for both, PED and ASME type safety
valves.
1. Remove the autoclave cover (see para. 9.2 “Dismantling the Outer
Covers of the Autoclave”).
2. Remove the water reservoir cover.
3. Unscrew the safety valve and remove it from the safety valve base.
4. Replace the valve with a new safety valve (install an original only!).
use Teflon tape for sealing the thread.
5. Perform one cycle and verify that the valve operates correctly.
ASME approved Type
CE marked Type