9.4 Replacing the Air Jet
Caution!
Before starting, be sure that the electric cord is disconnected and that
there is no pressure in the autoclave.
1. Remove the autoclave cover (see para. 9.2 Dismantling the Outer
Covers of the Autoclave).
2. Remove the water reservoir cover.
3. Remove the water reservoir silicon gasket.
4. Unscrew the air Jet with a 10 mm wrench and remove it from the
safety valve base.
5. Replace the valve with a new air Jet (install an original only!). use
Teflon tape for sealing the thread.
6. Test any autoclave cycle to verify that the valve operates correctly.
AIR JET