54
Corrections
2.1 See symptom s 29 in para. 11.3.2.
2.2 Increase the Program interval. Add 5-10
seconds to the water inlet.
3.1 Clean the piping or the valve.
Possible cause check-up and tests
1. Press door switch while the door is open and
simultaneously press START pus-button.
Message Water inlet will appear on the
display and water enters the chamber until it
reaches the grove adjacent to the door. On
this point the water must stop inserting the
chamber.
If the autoclave does not operate as described
above, proceed as follows:
2.1 Incorrect reading of the water sensing
electrode,
2.2 Programmed interval after the electrode
senses water surface is very short.
3.1 Probably water piping or valve is clogged.
This leads to a long filling time,
11.3 In Process Malfunction
Water Inlet Stage
Symptom
1. If there is any malfunction
related to inlet water
quqntity (Low Water, High
Temp, L;ow Press or Low
Heat), check the automatic
water filling.
2. The water inlet phase is
short and the water level in
the chamber is low.
3. The water inlet time is
prolonged and the quantity
of water is not acceptable.