55
Corrections
4.1 Relocate heat exchanger and release float
or clean the float.
4.2 Replace float.
4.3 Fix disconnection.
4.4 Replace plunger or complete solenoid.
4.5 Clean water inlet pipe
4.6 See symptom s 11 & 13 in para. 11.3.2.
Replace water inlet valve.
4.7 Replace plunger or complete solenoid.
4.8 Replace air release valve.
Possible cause check-up and tests
Operate the autoclave. Open the door while in
“water inlet” stage. If no water enters the
chamber, there is no water in the reservoir and
display indicates that there is water:
4.1 Float is stuck. Check if heat exchanger
moved and prevents the float from moving or
if float is stuck due to dirt.
4.2 Float is faulty.
There is water in the reservoir and the display
indicates that there is water. Despite this, water
does not enter the chamber.
4.3 No voltage at solenoid input.
Verify command at Tewst Point TP9 from
electronic box.
4.4 solenoid faulty.
4.5 Water inlet pipe is clogged.
4.6 Water inlet valve (21) is faulty.
4.7 Valve’s solenoid is faulty.
4.8 Air release valve (41) is faulty.
Symptom
4. Water does not enter
chamber during “FILL
WATER” stage.