Chapter 6 ___________________________________________________Calibration and Adjustment  
VAISALA______________________________________________________________________ 119 
6.  Enter the command MPCP1 ON (and MPCP2 ON, MPCP3 
ON, if applicable) to activate the new adjustments. 
7. Leave 
LCP1 (and LCP2, LCP3, if applicable) in the OFF state 
after multipoint adjustment. 
NOTE 
The new multipoint adjustments cancel the previous adjustments as 
well as the valid date of calibration of the transmitter. A new 
calibration date and note can be set with the CDATE and CTEXT 
commands. 
 
MPCP1 [x/y] [z] 
where: 
x = ON 
y = OFF 
z = ? 
 
Examples: 
Display current multipoint adjustments: 
>MPCP1 ? 
Multipoint adjustment: OFF 
Reading                   Reference                 Correction 
         800.21 hPa                 800.10 hPa                 -0.110 hPa 
         850.22 hPa                 849.20 hPa                 -1.020 hPa 
         901.02 hPa                 899.99 hPa                 -1.030 hPa 
         949.98 hPa                 950.01 hPa                  0.030 hPa 
        1000.11 hPa                1000.10 hPa                 -0.010 hPa 
        1050.12 hPa                1050.12 hPa                  0.000 hPa 
> 
 
Disable MPCP1 and LCP1 adjustments: 
>MPCP1 OFF 
Multipoint adjustment: OFF 
>LCP1 OFF 
Linear adjustment: OFF 
>