USER'S GUIDE____________________________________________________________________  
120 _________________________________________________________________ M210855EN-D 
Enter MPCP1 adjustment data, six points: 
>MPCP1 
1. Reading ? 800.21 
1. Reference ? 800.10 
2. Reading ? 850.22 
2. Reference ? 849.2 
3. Reading ? 901.02 
3. Reference ? 899.99 
4. Reading ? 949.98 
4. Reference ? 950.01 
5. Reading ? 1000.11 
5. Reference ? 1000.10 
6. Reading ? 1050.12 
6. Reference ? 1050.12 
7. Reading ? 
(Leave empty to finish data entry.) 
Multipoint adjustment: OFF 
Reading                   Reference                 Correction 
         800.21 hPa                 800.10 hPa                 -0.110 hPa 
         850.22 hPa                 849.20 hPa                 -1.020 hPa 
         901.02 hPa                 899.99 hPa                 -1.030 hPa 
         949.98 hPa                 950.01 hPa                  0.030 hPa 
        1000.11 hPa                1000.10 hPa                 -0.010 hPa 
        1050.12 hPa                1050.12 hPa                  0.000 hPa 
           0.00 hPa                   0.00 hPa                  0.000 hPa 
           0.00 hPa                   0.00 hPa                  0.000 hPa 
> 
 
Activate MPCP1 adjustment: 
>MPCP1 ON 
Multipoint adjustment: ON 
>