6.2 Changing the probe filter
New compatible filter
Clean lint-free gloves
Sensors are easily damaged when the filter is not in place. Handle
the probe head carefully.
CAUTION!
HMP9 and TMP1 probe models do not have a removable filter.
1. Put on clean gloves before touching the filter.
2. Turn the filter counter-clockwise to loosen it.
3. Remove the filter from the probe head. Be careful not to touch the sensors with the filter.
4. Install a new filter on the probe head. Tighten the filter properly (recommended force
5 Nm).
6.3
 Replacing the HUMICAP R2 sensor
New HUMICAPâ R2 sensor
New compatible filter
Clean lint-free gloves
11.3 %RH and 75.5 %RH humidity references (using Vaisala HMK15 Humidity
Calibrator)
Computer with Microsoft Windowsâ operating system and Vaisala Insight
software installed
USB connection cable (no. 242659)
Follow this procedure to replace a HUMICAPâ R2 humidity sensor it has been damaged, or
normal adjustment is not sucient to restore the humidity measurement accuracy. It is
recommended that you replace the filter at the same time. The procedure includes a special 2-
point adjustment of humidity measurement using Insight PC software. This adjustment must
be done every time after the HUMICAPâ R2 sensor has been replaced.
Chapter 6 – Maintenance
37