136 [5.35]
Ø 25
[0.98]
M12/5
41 [1.61]
253 [9.96]
243 [9.57]
Ø 12
[0.47]
Ø 13.5
[0.53]
Probe cable
2 m [6.56 ft] or 10 m [32.8 ft]
Ø 5
[0.2]
mm
[in]
Figure 32 HMP5 probe dimensions
8.4
HMP7 specifications
Table 18 Measurement performance
Property Description/value
Relative humidity
Measurement range 0 … 100 %RH
Accuracy at +23 °C (+73.4 °F)
1)
±0.8 %RH (0 … 90 %RH)
Factory calibration uncertainty
2)
±0.5 %RH (0 … 40 %RH)
±0.8 %RH (40 … 95 %RH)
T
63
response time 15 s
Sensor options HUMICAPâ R2
HUMICAPâ R2C
3)
HUMICAPâ 180VC
3)
4)
Temperature
Measurement range −70 … +180 °C (−94 … +356 °F)
Accuracy at +23 °C (+73.4 °F)
1)
5)
±0.1 °C (±0.18 °F)
HMP3, HMP4, HMP5, HMP7, HMP8, HMP9, MMP8, TMP1 User Guide M212022EN-D