Property Description/value
Factory calibration uncertainty
2)
±0.1 °C (±0.18 °F) at +23 °C (+73.4 °F)
Sensor Pt100 RTD Class F0.1 IEC 60751
1) Defined against calibration reference. Including non-linearity, hysteresis, and repeatability.
2) Defined as ±2 standard deviation limits. Small variations possible; see calibration certificate.
3) Chemical purge feature available with this sensor
4) Exposing temperature sensor to temperatures outside −20 … +150 °C (−4 … +302 °F) may
cause permanent additional deviation of ±0.1 °C (0.18 °F)
0.0
0.5
1.0
1.5
2.0
2.5
3.0
-40
-30
-20
-10
0
10
20
30
40
50
60
70
80
90
100
110
120
Accuracy [+/- % RH]
Temperature [
90 ... 100 %RH
0 ... 90 %RH
Figure 36 HMP8 humidity measurement accuracy as a function of temperature
Accuracy [+/-
°C]
Temperature [
-40
-30
-20
-10
0
10
20
30
40
50
60
70
80
90
100
110
120
130
140
150
160
170
180
Figure 37 HMP8 temperature measurement accuracy over full range
HMP3, HMP4, HMP5, HMP7, HMP8, HMP9, MMP8, TMP1 User Guide M212022EN-D
62