Property Description/value
Accuracy at +23 °C (+73.4 °F)
1)
±0.8 %RH (0 … 90 %RH)
Factory calibration uncertainty
2)
±0.7 %RH (0 … 40 %RH)
±1 %RH (40 … 95 %RH)
Non-linearity 0.4 %RH
T
63
response time
3)
15 s
Sensor HUMICAPâ I
Temperature
Measurement range −40 … +120 °C (−40 … +248 °F)
Accuracy at +23 °C (+73.4 °F)
1)
±0.1 °C (±0.18 °F)
Factory calibration uncertainty
2)
±0.1 °C (±0.18 °F) at +23 °C (+73.4 °F)
T
63
response time
3)
70 s
1) Defined against calibration reference. Including non-linearity, hysteresis, and repeatability.
2) Defined as ±2 standard deviation limits. Small variations possible; see calibration certificate.
3) In still air
Table 27 Operating environment
Property Description/value
Operating temperature of probe body −40 … +60 °C (−40 … +140 °F)
Operating temperature of probe head −40 … +120 °C (−40 … +248 °F)
Storage temperature −40 … +60 °C (−40 … +140 °F)
Operating environment Suitable for outdoor use when protected from
rain
Measurement environment For air, nitrogen, hydrogen, argon, helium, and
oxygen
1)
IP rating IP65 for probe body
EMC compliance EN 61326-1, industrial electromagnetic
environment
1) Consult Vaisala if other chemicals are present. Consider safety regulations with flammable
gases.
Chapter 8 – Technical data
65
陕西威瑞仪器仪表有限公司 生产代理销售:实验检测设备 气象科学仪器设备
Tel:029-88186182 Web:www.xavery.cn Email:sxvery@163.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Vaisala HMP3 and is the answer not in the manual?

Vaisala HMP3 Specifications

General IconGeneral
BrandVaisala
ModelHMP3
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals