Vaisala HMP3 - Figure 24 HMP3 Humidity Measurement Accuracy as a Function of Temperature; Figure 25 HMP3 Temperature Measurement Accuracy over Full Range; Table 7 Operating Environment

Vaisala HMP3
96 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
0.0
0.5
1.0
1.5
2.0
2.5
3.0
-40
-30
-20
-10
0
10
20
30
40
50
60
70
80
90
100
110
120
Accuracy [+/- % RH]
Temperature [
°
C]
3.5
4.0
90 ... 100 %RH
0 ... 90 %RH
Figure 24 HMP3 humidity measurement accuracy as a function of temperature
0
0.1
0.2
0.3
0.4
Accuracy [+/-
°C]
Temperature [
°
C]
-40
-30
-20
-10
0
10
20
30
40
50
60
70
80
90
100
110
120
Figure 25 HMP3 temperature measurement accuracy over full range
Table 7 Operating environment
Property Description/value
Operating temperature of probe head −40 … +120 °C (−40 … +248 °F)
Operating temperature of probe body −40 … +80 °C (−40 … +176 °F)
Storage temperature −40 … +80 °C (−40 … +176 °F)
Operating environment Suitable for outdoor use
Measurement environment For air, nitrogen, hydrogen, argon, helium, and
oxygen
1)
IP rating of probe body IP66
EMC compliance EN61326-1, industrial environment
1) Consult Vaisala if other chemicals are present. Consider safety regulations with flammable
gases.
HMP3, HMP4, HMP5, HMP7, HMP8, HMP9, MMP8, TMP1 User Guide M212022EN-D
50