Vaisala Indigo 201 - Figure 22 Start Calibration Button; Starting and Closing Calibration Mode

Vaisala Indigo 201
84 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
the transmitter is a new unit and has not been configured yet, or
you have removed the analog output configuration from Indigo manually.
It is recommended to always check the analog output settings from the wireless
configuration interface's Settings > Outputs menu after connecting a probe.
More Information
Receiving Analog Output Settings from Probe (page 16)
Clearing Analog Output Settings from Indigo (page 47)
5.2 Analog Output Configuration
Example
This example shows how to configure analog output 1 to use the following settings when a
CO
2
probe is connected to Indigo:
4 ... 20 mA current output
Carbon dioxide measurement scaled to 0 ... 20 %CO
2
Clipping limit 5%, error limit 10%, and error output 3.6 mA
When you enter a value into a
field, the value is saved automatically when you exit the
input field (for example, tap on an area outside of the field).
To configure the output:
1. Launch the Indigo 201 wireless
configuration interface.
2. Open the Settings > Outputs menu.
3. Select Current outputs from the Mode dropdown.
4. Select 4 ... 20 mA from the Output Type dropdown.
5. Select Carbon dioxide concentration from the Parameter dropdown.
6. Select % from the Unit dropdown.
7. Type 0 in the Scale Low End
field.
8. Type 20 in the Scale High End field.
9. Type 5 in the Clipping Limit field and 10 in the Error Limit field.
10. Type 3.6 in the Error Output Current field.
More Information
Output Settings (page 39)
Indigo 201 User Guide M211877EN-E
46