Table 5 Indigo Transmitter Messages Shown on Display
Display Message Description Recommended Action
Errors
Low supply voltage The supply voltage is below the
minimum (range: 15 ... 30 VDC or
20 ... 22 VAC).
The error clears when the supply
voltage returns to the specified
range. Check your power supply
and restart Indigo if needed.
No legal measurement probe The connected probe is not
compatible with Indigo or a
restriction (probe name or serial
number) on allowed probes is in
place.
Check your probe's Indigo
compatibility and possible
restrictions on allowed probes.
Analog output 1 out of range The output of analog output 1 has
gone out of range.
The error clears when the output
returns to range. Check the analog
output configuration.
Analog output 2 out of range See above See above
Analog output 3 out of range See above See above
WLAN error There is a problem with the
wireless connection.
Go through the troubleshooting
instructions in Indigo Wireless
Connection Troubleshooting
(page 65).
Warnings
Outputs configured for dierent
probe
An analog output configuration
already exists in Indigo, and the
settings cannot be received
automatically from the probe you
are connecting.
Configure the analog output
settings for the new probe
manually, or reset the analog
output settings in Indigo to enable
automatic receiving of analog
output settings.
No measurement probe The measurement probe is not
connected, or the probe is
connected poorly and cannot be
recognized by Indigo.
Check that you have fastened the
probe correctly with the locking
wheel.
Notifications
Probe connected: Indigo displays the name of the
connected probe.
WLAN on The WLAN connection is active
and you can connect to the
wireless configuration interface.
Waiting for measurements The connected probe is preparing
measurements (for example, at
start-up or after a probe restart).
Chapter 8 – Maintenance and Troubleshooting
69