Table 5 Indigo Transmitter Messages Shown on Display
Error Message Description Recommended Action
Errors
Low supply voltage The supply voltage is below the
minimum (r
ange: 15 ... 30 VDC or
20 ... 22 VAC).
The error clears when the supply
voltage returns to the specified
range. Check your power supply
and restart Indigo if needed.
No legal measurement probe The connected probe is not
c
ompatible with Indigo or a
restriction (probe name or serial
number) on allowed probes is in
place.
Check your probe's Indigo
compatibility and possible
restrictions on allowed probes.
WLAN error There is a problem with the
wir
eless connection.
Go through the troubleshooting
instructions in Indigo Wireless
Connection Troubleshooting.
Warnings
No measurement probe The measurement probe is not
c
onnected, or the probe is
connected poorly and cannot be
recognized by Indigo.
Check that you have fastened the
probe correctly with the locking
wheel.
Notifications
Probe connected: Indigo displays the name of the
c
onnected probe.
WLAN on The WLAN connection is active
and y
ou can connect to the
wireless configuration interface.
Chapter 8 – Maintenance and Troubleshooting
59