EasyManua.ls Logo

Vaisala Indigo 202 - Input;Output Specifications

Vaisala Indigo 202
76 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3.5 Input and Output Specification
Table 2 Input and Output
Property Specification
Digital output RS-485 Modbus RTU
Relays 2 configurable relays (VAC/VDC)
Device maximum specification (resistive load):
Max. switching power 30 W / 37.5 VA
UL-rated maximum
specification (resistive load):
AC: max. 28 V / 0.5 A
DC: max. 40 V / 0.24 A
Up to 30 VDC:
max. switching current 1 A
max. switching power 30 W
Power supply input
1)
Range 15 ... 30 VDC (20 ... 22 VAC)
Maximum current Transmitter and connected probe max. 1 A
Power consumption Transmitter max. 3 W (+ connected probe, varies
depending on pr
obe type)
Probe connector M12/5 connector for probe or probe cable connection
(V
aisala Indigo-compatible probes)
Cable feed throughs 2 options: rubber lead-through on the bottom of the
tr
ansmitter, and opening with a seal at the back of the
transmitter
Screw terminal wire size
0.2 mm
2
... 1.5 mm
2
1) Using a power supply with overload protection is recommended for electrical safety.
Do not modify the unit or use it in ways not described in the documentation.
Impr
oper modification may lead to safety hazards, equipment damage, failure to
perform according to specification, or decreased equipment lifetime.
CAUTION!
More Information
Specifications
(page 61)
Chapter 3 – Installation
17

Table of Contents

Other manuals for Vaisala Indigo 202

Related product manuals