EasyManua.ls Logo

Vaisala Indigo 520 - Relay Output Terminals and Lead-Through; Figure 9 Example of Stripped AC (Mains) Power Cable; Table 4 AC Power Supply Input Terminals

Vaisala Indigo 520
82 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Table 4 AC power supply input terminals
Terminal Function Notes
Power supply ground
L Line 100 … 240 VAC 50/60 Hz
N Neutral
3.3.2.1 Preparing AC (mains) power cable
Cable stripping tool
Prepare the AC (mains) power cable for connection to the power supply input terminal:
mm
[in]
55 [2.17]
45 [1.77]
8 [0.31]
1
2
3
Figure 9 Example of stripped AC (mains)
power cable
1 Line wire
2 Grounding wire
3 Neutral wire
1. Strip 55 mm (2.17 in) of the cable to expose the wires.
2. Cut o 10 mm (0.39 in) of the line and neutral wires. Leave the grounding wire 55 mm
(2.17 in) long.
Make sure that the grounding wire is longer than the line and
neutral wires. Under mechanical stress, the grounding wire must be the last
to disconnect from the protective ground terminal.
CAUTION!
3. Strip the ends of the individual wires to expose the conductors for a length of 8 mm
(0.32 in).
3.3.3 Relay output terminals and lead-through
Before connecting wires or cables, make sure that the transmitter is powered o.
Chapter 3 – Installation
25

Table of Contents

Other manuals for Vaisala Indigo 520

Related product manuals