EasyManua.ls Logo

Vaisala Indigo 520 - Probe Connection Terminals and Lead-Throughs; Figure 10 M20×1.5 Cable Gland with Split Bushing

Vaisala Indigo 520
82 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1
2
3
4
Figure 10 M20×1.5 cable gland with split
bushing
1 Base of the cable gland
2 Nylon seal
3 Split bushing inside the seal
4 Nut of the cable gland
3.3.6 Probe connection terminals and lead-throughs
Before connecting wires or cables, make sure that the transmitter is powered o.
ANALOG OUTPUTS
RELAY 1RELAY 2
+
+
+
+
+
+
+
+
+
CH1
CH2
CH3
CH4
RS-485
PROBE
24 VOUT
ANALOG
INPUT
NO COM NC NO COM NC
1
2
VS
+
1
2
For the M16×1.5 cable glands ordered
together with the transmitter from Vaisala,
the cable diameter is 2.0 … 6.0 mm
(0.08 … 0.24 in). Tightening torque for the
cable gland is 6 Nm.
The recommended maximum length of the
probe cable is 30 m (98 ft).
1 Probe 1 lead-through, M16×1.5
2 Probe 2 lead-through, M16×1.5
Indigo 520 User Guide M212287EN-A