For the M20×1.5 cable gland with split
bushing, the cable diameter is 7 mm
(0.28 in).
Table 6 Analog output terminals
Terminal Function Notes
CH1 + Analog output channel 1 +
Max. wire size: 2.5 mm
2
(14 AWG)
CH1 - Analog output channel 1 -
CH2 + Analog output channel 2 +
CH2 - Analog output channel 2 -
CH3 + Analog output channel 3 +
CH3 - Analog output channel 3 -
CH4 + Analog output channel 4 +
CH4 - Analog output channel 4 -
3.3.5 Ethernet connector and lead-through
Before connecting wires or cables, make sure that the transmitter is powered o.
The 8P8C (RJ45) Ethernet connector is located on the inside of the transmitter cover. The
supported standards are 10BASE-T and 100BASE-TX.
ANALOG OUTPUTS
RELAY 1RELAY 2
+
+
+
+
+
+
+
+
+
CH1
CH2
CH3
CH4
RS-485
PROBE
24 VOUT
ANALOG
INPUT
NO COM NC NO COM NC
1
2
SERVICE PORT
PRESS TO ACTIVATE
WEB INTERFACE
ETHERNET
M20×1.5
VS
+
For the M20×1.5 cable gland with split bushing, the cable diameter is 7 mm (0.28 in).
When you insert the cable through the gland parts, also remove the split bushing
(3) inside the nylon seal (2) to make the cable fit through the seal. Then place the
bushing around the cable and push it back inside the seal. See the following
figure.
Chapter 3 – Installation
27