Problem: The web interface does not show the same contents as the touchscreen display.
Possible cause: Solution:
You have changed a setting on
the touchscreen, but the
change is not shown in the web
interface before you refresh the
web page.
Refresh the web page.
Problem: There is no measurement data from the connected probe.
Possible cause: Solution:
The probe may be
disconnected from the
transmitter.
1. If the connected probe has a status indicator, check its color.
2. If the indicator is o, make sure that the probe is properly
connected to the probe cable connector.
3. If the indicator is red, check the probe sensor. If the sensor is
faulty, replace it.
4. If the indicator is blinking green, restart the transmitter.
5. If the problem still persists, contact Vaisala.
The probe sensor may be
faulty.
More information
Opening and closing transmitter cover (page 17)
Connecting to web interface (page 39)
Creating web interface users (page 40)
Logging in to web interface (page 42)
Restoring factory default settings (page 66)
Chapter 9 – Maintenance and troubleshooting