Overview of calibration and adjustment with Indigo80 (page 23)
Configuring probe features with Indigo80 handheld indicator (page 31)
Configuring environment settings with Indigo80 handheld indicator (page 32)
3.5.1 Overview of calibration and adjustment with Indigo80
The Indigo80 handheld indicator provides two ways to calibrate and adjust measurements for
probes: Guided calibration and Manual adjustment.
Figure 10 Indigo80 calibration methods for
probes
Choose Guided calibration when you want the indicator to guide you through the
calibration process step by step. At the end you can also adjust the measurement, if
needed.
Choose Manual adjustment when you want to freely set the reference and measured
values to adjust the measurement.
Calibrating a measurement with the indicator, including purge and adjustment, typically takes
1 ... 2 hours depending on the number of calibration points.
For detailed information on using Indigo80 for calibration and adjustment, see
your Indigo80-compatible probe's user documentation at docs.vaisala.com.
If you have a transmitter connected to Indigo80, you can also use the indicator to
test and adjust the input and output channels of the transmitter. For more
information, see your transmitter's user documentation at docs.vaisala.com.