Acknowledgement
(within 10ms after reception of command)
This command initiates ZERO to compensate for offset of gauge(s).
Note: Refer to «ZERO» for correct zero procedure.
[c:6002][xxxxxxxx][CR][LF]
data length: 8 characters
xxxxxxxx System base pressure, value depends on configuration,
refer to «RS232 setup commands, RANGE CONFIGURATION»
for details. Alignment range is equivalent to max. +/-1.4V sensor signal.
This command aligns PRESSURE to a certain value. Also SENSOR READING will be
aligned accordingly. It might be used instead of ZERO in case base pressure is not low
enough.
data length 8 characters starting with a zero
xxxxxxx Pressure limit for LEARN, value depends on configuration,
refer to «RS232 setup commands, RANGE CONFIGURATION»
for details
This command starts LEARN.
By OPEN VALVE, CLOSE VALVE or POSITION CONTROL commands the routine may
be interrupted.
Note: Without LEARN the PID controller is not able to perform pressure control. Refer to
«LEARN» for correct learn gas flow and procedure.
[d:][pppdddddddd][CR][LF]
data length 3 + 8 characters
ppp pointer, 000 ... 103
dddddddd single data set
This command downloads the LEARN data sets from the host computer to the valve.
There are a total number of 104 data sets. Each data set consists of 8 data bytes and
needs to be uploaded separately.
Note: Make sure that all 104 data sets will be downloaded.
[u:][pppdddddddd][CR][LF]
data length 3 + 8 characters
ppp pointer, 000 ... 103
dddddddd single data set
This command uploads the LEARN data sets from the valve up to the host. There are a
total number of 104 data sets. Each data set consists of 8 data bytes and needs to be
uploaded separately.
Note: Make sure that all 104 data sets will be uploaded.