INSTALLATION  Series 
 
32/129
Edition 2019-05-03  963283EA
 
4.3.2.4  Power and sensor connection (
±
15 VDC sensors) without opt. SPS module external 
[653 . . - . . G . - . . . . / 653 . . - . . H . - . . . . versions recommended] 
 
 
 
 
 
• 
VAT fuse recommendation: (a) 5 AF 
• 
Use shielded sensor cable(s). Keep cable as short as possible, but locate it away 
from noise sources. 
• 
Connect Power supply (+24 / GND) at DA–15 male power connector and Sensors 
(0V / + / -) at DA–15 female sensor connector exactly as shown in the drawing 
above! 
• 
Connector: Use only screws with 4–40 UNC thread for fastening the connectors!
 
 
 
 
Pins 4 and 11 must be 
bridged for operation. 
An optional switch 
would allow for motor 
interlock to prevent 
valve from moving.
 
 
Low range sensor may be 
connected to sensor 1 or 
sensor 2 input. Do 
configuration accordingly.