Series 65  INSTALLATION 
 
963283EA  Edition 2019-05-03  47/
 
 
 
4.8.2  Zero Adjust 
Zero Adjust allows for the compensation of the sensor offset voltage. 
When Zero Adjust is performed the current value at the sensor input is equated to Zero.Adjust Target 
Pressure. 
In case of a 2 sensor system both sensor inputs will be adjusted. 
Note:  A maximum offset voltage of +/- 1.4 V can be compensated. 
  To execute a zero adjust, the zero adjust have to be enabled (see chapter above). 
 
Parameter  Description 
Selection 
Select the sensor for the zero adjust: 
•  Sensor 1 + 2 
•  Sensor 1 
•  Sensor 2 
Pressure 
Normally this parameter is set to 0 in case the process chamber is fully 
evacuated (which means pressure value is nearly 0). If not you can align 
the sensor value to a known pressure. In this case set Target Pressure 
to the current pressure. 
Note: Target Pressure is in the unit of pressure, see chapter «Scaling of 
Pressure and Position Values» 
Start the zero adjust by set 
 to 1 
At this moment offset value will be calculated and Sensor x.Value = 
Target Pressure. The resulting offset value can be read on parameter 
Sensor x.Offset Value [SFS] 
 
Clear the offset values by setting Execute to 2 
The result of clearing the zero adjust: Sensor x.Offset Value [SFS] = 0.0 
Sensor 2.Enable 
0:  It’s not possible to execute a zero adjust. A present offset 
value is ignored. 
1:  It’s possible to execute a zero adjust. A present offset value 
[SFS] 
Sensor 2.Offset Value 
[SFS] 
Value which will be subtracted from measured sensor value 
The value is related to sensor full scale (0.1 means 10% of sensor full 
scale) 
 
 
Parameter location: 
CPA  RS232 
Pressure Sensor->Zero Adjust 
Pressure Sensor->Sensor 1->Zero Adjust 
Pressure Sensor->Sensor 2->Zero Adjust 
Pressure Sensor->Sensor 1->Zero Adjust 
Pressure Sensor->Sensor 2->Zero Adjust 
Refer to “RS232 setup commands” for 
details. 
 
Performing a zero adjust: 
 
1.  Wait until process chamber is evacuated and sensor signal is not shifting anymore. Do not 
perform Zero Adjust as long as pressure gauge voltage is shifting otherwise incorrect pressure 
reading is the result. Refer to manual of sensor manufacturer for warm up time. 
2.  Set parameter Sensor Selection 
3.  Set parameter Target Pressure (Zero Adjust->Target Pressure, not the Target Pressure for 
pressure control) 
4.  Set parameter Execute = 1 
5.  Check parameter Actual Pressure if the pressure is shifted as expected