4. Installing the X-ray Unit: Base Stand Type
Green X / X18
(Model: PHT-75CHS) Installation Manual 63
4.7.1 Leveling Right and Left
To level the equipment right and left:
To obtain the accurate level, place the spirit level
as illustrated.
1. Place the spirit level inside the vertical unit, as shown in the illustration below.
2. Use a T-shaped hex wrench to adjust the screws in circled locations until the
bubble inside spirit level comes to the middle.