EasyManuals Logo
Home>Vatech>Dental equipment>PaX-i3D Smart

Vatech PaX-i3D Smart User Manual

Vatech PaX-i3D Smart
156 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #140 background imageLoading...
Page #140 background image
13. Appendices
133
I
PaX-i3D Smart
English
CARPUS
Figure
Gender
Hard (kVp / mA) Normal (kVp / mA) Soft (kVp / mA)
Man
60/6 60/5 60/4
Woman
60/6 60/5 60/4
Child
60/6 60/5 60/4
* Tube voltage and current variation range in the CEPHALOMETRIC mode (recommended)
: 60 ~ 90 kVp, 4 ~ 12 mA
C. CBCT
Mode
Low Dose Ultra Low Dose
kVp mA kVp mA
Man
Hard
95 6.6 80 5.0
Normal
94 6.6 79 5.0
Soft
93 6.6 78 5.0
Woman
Hard
95 6.3 80 4.7
Normal
94 6.3 79 4.7
Soft
93 6.3 78 4.7
Child
Hard
95 6.3 80 4.7
Normal
94 6.3 79 4.7
Soft
93 6.3 78 4.7