X-ray Scatter Dose
Test Equipment Information
Instrument Manufacturer Model S/N
Radiation Monitor
Controller
Radcal 9015 91-1470/19069
PANO Mode
Mode Tested: PANO Normal Adult 13.6s
Mode Tested
PANO HD
Applied Tube Voltage Peak [kVp]
80
Applied Tube Current [mA]
14
Mode
Direction [°]
PANO Normal Adult 13.6s [mR/hr]
1 m (3.3 ft) 1.5 m (4.9 ft) 2 m (6.6 ft)
0 60.9 17.7 8
45 19.6 12.4 5.8
90 10.6 6.8 4.1
135 22.1 12.5 5.6
180 1 0 0
225 45.4 21.4 9.4
270 47.6 21.9 9.2
315 76.4 19.4 8.6