EasyManua.ls Logo

Vatech PaX-i3D Smart

Vatech PaX-i3D Smart
156 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
13. Appendices
138
I
PaX-i3D Smart
CT Mode
Test Condition
Tested Mode
CT FOV 100 x 85 18s Low Dose
Applied Tube Voltage Peak [kVp]
99
Applied Tube Current [mA]
14
Mode
Direction [°]
Ultra Low Dose, FOV 100 x 80 5.9s [mR/hr]
1 m (3.3 ft) 1.5 m (4.9 ft) 2 m (6.6 ft)
0 588.2 135.3 87.1
45 549.3 249.4 106.8
90 472.6 307.3 78.4
135 458.8 287.6 89.3
180 12.9 4.6 1.3
225 410.5 288.7 98.2
270 663.2 301.4 112.4
315 429.7 194.2 92.3

Table of Contents

Other manuals for Vatech PaX-i3D Smart

Related product manuals