PaX-i3D Green (PHT-60CFO) User Manual
141
12 Technical Specications
Item Description
X-ray
tube
Focal spot size 0.5 mm (IEC60336)
Target angle 5 °
Heat storage capacity
(Maximum heat content)
35 kJ 30 kJ
Inherent ltration
At least 0.8 mm
Al eq.at 50 kV
0.5 mm Al
X- ray coverage 95 x 380 mm at SID 550 mm
Anode heat content 35 kJ _
Normal anode input
power
1,750 W (at 1.0 s)
2,000 W
(at 1.0 s)
Maximum anode heat
dissipation rate
250 W
Duty cycle
1:60 or more
(Exposure time : interval time)
D-052SB
Maximum Rating Charts
DC (Center Grounded)
[PI3DG_130U_44A_en]User Guide.indd 141 2016-05-24 오후 4:19:21