PaX-i3D Green (PHT-60CFO) User Manual
150
Appendices
2 X-ray Dose Data
The X-ray dose data is extracted from the X-ray Dose Test Report for the
PaX-i3D Green.
X-ray Dose Test Report for the PaX-i3D Green maintains dosemetric evaluation
of VATECH dental diagnostic system meets all requirements specied in the
IEC Collateral Standard. To limit unnecessary exposure to the patient, operator
or other staff, the PaX-i3D Green is designed to comply with IEC 60601-1-3
Part 1 General Requirements for Safety.
Test Condition
Brand Name (Model)
PaX-i3D Green (PHT-60CF0)
Sensor type
PANO: Xmaru1501CF
CEPH: 1210SGA (One shot Type)
CT: Xmaru1215CF Master Plus
Xmaru1524CF Master Plus
X-ray generator
DG-07C11T2
2.1
DAP Table
Test Equipment
Instrument Manufacturer Model S/N
Dose Meter Piranha 205 CB2-06090029
Mode Tested: PANO HD Normal Adult Standard 13.5s
mA
kVp
6 8 10
[mGy] [mGy
ㆍ
cm2] [mGy] [mGy
ㆍ
cm2] [mGy] [mGy
ㆍ
cm2]
60
5.1 36.5 6.7 48.4 8.2 59.2
70
6.6 47.3 8.7 62.7 10.7 76.7
80
8.5 61.4 11.3 81.4 13.8 99.6
90
10.5 75.5 13.9 100.0 17.0 122.3
[PI3DG_130U_44A_en]User Guide.indd 150 2016-05-24 오후 4:19:22