PaX-i3D Green (PHT-60CFO) User Manual
149
Appendices
[CARPUS]
Figure Hard (kVp / mA) Normal (kVp / mA) Soft (kVp / mA)
One shot Scan
Fast
Scan
One shot Scan
Fast
Scan
One shot Scan
Fast
Scan
Man
60/6 60/6 60/10 60/5 60/5 60/9 60/4 60/4 60/8
Woman
60/6 60/6 60/10 60/5 60/5 60/9 60/4 60/4 60/8
Child
60/6 60/6 60/10 60/5 60/5 60/9 60/4 60/4 60/8
* Tube voltage and current variation range in the CEPHALOMETRIC mode: 60 ~ 100 kVp, 4 ~ 14 mA
C. CBCT
Xmaru1215CF master plus / Xmaru 1524CF master plus
Mode
High Resolution Green
kVp mA kVp mA
Man
Hard 95 10.0 95 10.0
Normal 95 9.0 95 9.0
Soft 95 8.5 95 8.5
Woman
Hard 95 9.5 95 9.5
Normal 95 8.5 95 8.5
Soft 95 8.0 95 8.0
Child
Hard 95 9.2 95 9.2
Normal 95 8.2 95 8.2
Soft 95 7.7 95 7.7
Depending on the circumference of the patient’s head, X-ray intensity may be
classied as Hard, Normal, or Soft :
Soft ≤ Normal ≤ Hard
Age Group
Average head
circumference
Range (cm)
classication of head
circumference
Child
53
±
3 cm
> 53
±
3 Hard
53
±
3 Normal
53
±
3 < Soft
Adult
56
±
3 cm
> 56
±
3 Hard
56
±
3 Normal
56
±
3 < Soft
- Maximally allowed tube voltage / current: kVp ± 10 % / mA ± 20 % according to
IEC60601-2-7.
- Due to image optimization performed prior to shipping, equipment data may differ
slightly from those specied in the table.
Gender
Sensor
Type
[PI3DG_130U_44A_en]User Guide.indd 149 2016-05-24 오후 4:19:22