148
Appendices
1 Recommended X-ray Exposure Table
A. PANO
[Standard / TMJ / Sinus / CARPUS]
Gender / Figure Hard (kVp / mA) Normal (kVp / mA) Soft (kVp / mA)
Man
75 / 14 74 / 14 73 / 14
Woman
74 / 14 73 / 14 72 / 14
Child
68 / 10 67 / 10 66 / 10
[Fast Mode]
Gender / Figure Hard (kVp / mA) Normal (kVp / mA) Soft (kVp / mA)
Man
75 / 16 74 / 16 73 / 16
Woman
74 / 16 73 / 16 72 / 16
Child
68 / 16 67 / 16 66 / 16
* Tube voltage and current variation range in the PANORAMIC mode: 50 ~ 80 kVp, 4 ~ 16 mA
B. CEPH
[Lateral]
Figure Hard (kVp / mA) Normal (kVp / mA) Soft (kVp / mA)
One shot Scan
Fast
Scan
One shot Scan
Fast
Scan
One shot Scan
Fast
Scan
Man
92/14 87/10 97/14 90/14 85/10 95/14 88/14 83/10 93/14
Woman
90/14 85/10 95/14 88/14 83/10 93/14 86/14 81/10 91/14
Child
86/10 82/10 87/12 84/10 80/10 85/12 82/10 78/10 83/12
[PA / SMV / Waters View]
Figure Hard (kVp / mA) Normal (kVp / mA) Soft (kVp / mA)
One shot Scan
Fast
Scan
One shot Scan
Fast
Scan
One shot Scan
Fast
Scan
Man
94/14 90/10 99/14 92/14 88/10 97/14 90/14 86/10 95/14
Woman
92/14 88/10 97/14 90/14 86/10 95/14 88/14 84/10 93/14
Child
90/10 85/10 89/12 88/10 83/10 87/12 86/10 80/10 85/12