50
0900‑5005 REV‑00
REPROCESSING
Read the “Warnings & Cautions section before performing the following tasks.
Cleaning and disinfecting the BladderScan i10 system is an important part of using and maintainingit.
Before each use, ensure all system components have been cleaned and disinfected according to the guidance
provided in Table 10.
Note: It is understood that all items in the following table will be used as intended.
Table 10. Reprocessing Requirements for the BladderScan i10 System
COMPONENT
REQUIRED REPROCESSING LEVELS
Clean Low High
Console
Probe Cable
Probe
Printer
Workstation
The reprocessing levels shown in this table refer to the CDC/Spaulding classifications.
CLEANERS & DISINFECTANTS
DISINFECTION EFFICACY
Table 11 lists the low‑level disinfectant (LLD) classes that have demonstrated efficacy with the system. For
information about the long‑term effect of solutions on system materials, see “Compatibility”.
Table 11. Validated Disinfectants
ACTIVE INGREDIENT CLASS TESTED ACTIVE INGREDIENTS
Alcohols (or alcohol with Quat) 55% isopropyl alcohol
Chlorine & chlorine compounds 0.55% sodium hypochlorite (bleach)
Quaternary ammonium (Quat) 0.5% quaternary ammonium compounds
Hydrogen peroxide 2.0% hydrogen peroxide

Table of Contents

Other manuals for Verathon BladderScan i10

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Verathon BladderScan i10 and is the answer not in the manual?

Verathon BladderScan i10 Specifications

General IconGeneral
BrandVerathon
ModelBladderScan i10
CategoryMedical Equipment
LanguageEnglish

Related product manuals