RECOMMENDED SEPARATION DISTANCES
Table 28. Recommended Separation Distances between Portable and Mobile RF Communications
Equipment and the System
The system is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the system can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters)
and the system as recommended below, according to the maximum output power of the communications
equipment.
RATED MAXIMUM
OUTPUT POWER OF
TRANSMITTER (W)
SEPARATION DISTANCE ACCORDING TO FREQUENCY OF TRANSMITTER (m)
150kHz to 80MHz
d=1.2P
80MHz to 800MHz
d=1.2P
800MHz to 2.5GHz
d=2.3P
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d
in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is
the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note: At 80MHz and 800MHz, the separation distance for the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people.
ACCESSORY CONFORMANCE TO STANDARDS
To maintain electromagnetic interference (EMI) within certified limits, the system must be used with the
cables, components, and accessories specified or supplied by Verathon. For additional information, see the
Components & Accessories section on page10 and the Product Specifications section on page74.
The use of accessories or cables other than those specified or supplied may result in increased emissions or
decreased immunity of the system.
Table 29. EMC Standards for Accessories
ACCESSORY LENGTH
Probe cable 1.8m (6.0ft)
USB printer cable 1.7m (5.6ft)
AC power cord 4.5m (15.0ft)
Power adapter cable 1.5m (4.9ft)

Table of Contents

Other manuals for Verathon BladderScan i10

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Verathon BladderScan i10 and is the answer not in the manual?

Verathon BladderScan i10 Specifications

General IconGeneral
BrandVerathon
ModelBladderScan i10
CategoryMedical Equipment
LanguageEnglish

Related product manuals