83
Operations & Maintenance Manual: Product Specifications
0900‑5005 REV‑00
Table 27. Guidance and Manufacturer’s Declaration Electromagnetic Immunity
The system is intended for use in the electromagnetic environment specified below. The customer or the user
of the system should ensure that it is used in such an environment.
IMMUNITY TESTS IEC 60601 TEST LEVEL
COMPLIANCE
LEVEL
ELECTROMAGNETIC ENVIRONMENT
GUIDANCE
Radiated RF
IEC 61000‑43
3 V/m
80 MHz to 2.7 GHz
80% AM at 1 kHz
In compliance
Interference may occur in the vicinity
of equipment marked with the
following symbol:
Note: Ut is the AC mains voltage prior to application of the test level.
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.