16
SETTING UP
To reduce the risk of electrical shock, use only the accessories and peripherals recommended by
Verathon
®
.
WARNING
Before you can use the system for the first time, you must inspect the components, set up the system, and
perform a functional test as recommended by Verathon
®
. Complete the following procedures:
1. Perform Initial InspectionInspect the system for any obvious physical damage that may have occurred
during shipment.
2. Mount the GlideScope Video Monitor (Optional)—Set up the GlideScope Video Monitor on a mobile
stand or IV pole.
3. Charge the Monitor BatteryYou can use the system while the battery is charging.
Note: The monitor will operate without charging the battery by using the GlideScope Video Monitor 12V
DC Power Adapter that shipped with the unit.
4. Connect the Video Cable or Smart Cable to the MonitorAttach the cable that connects the video
baton or laryngoscope to the monitor and transmits the video data.
5. Connect the Smart Cable to the Video Baton (VideoBaton2.0Only)Attach the Smart Cable to a
GlideScope Video Baton 2.0.
6. Connect to an External Monitor (Optional)Connect the monitor to an external display source, such as a
larger monitor screen, by using the HDMI‑to‑DVI cable.
7. Configure User SettingsEnter data customized to your clinic, and configure settings such as the date
and time.
8. Perform a Functional CheckBefore you use the device for the first time, perform a functional check to
ensure that the system is working properly.
PROCEdURE 1. PERFORM INITIAL INSPECTION
When you receive the system, Verathon recommends that an operator familiar with the instrument perform a
full visual inspection of the system for any obvious physical damage that may have occurred during shipment.
1. Verify that you have received the appropriate components for your system by referring to the packing list
included with the system.
2. Inspect the components for damage.
3. If any of the components are missing or damaged, notify the carrier and Verathon Customer Care or your
local representative. For contact information, visit verathon.com/support.