26
PROCEdURE 9. PERFORM A FUNCTIONAL CHECK
Before you use the device for the first time, perform the following functional check to ensure that the system
is working properly. Please contact your local Verathon
®
representative or Verathon Customer Care if your
system does not function as described below. For contact information, visit verathon.com/support.
REQUIREd CHECKS
1. Fully charge the monitor battery (this takes approximately 6hours).
2. Attach the video cable to the monitor.
3. Press the Power button. The monitor turns on.
4. Look at the monitor screen, and verify that the image displayed is being received from the video baton or
GlideScope Direct blade.
Note: There may be a slight blade intrusion in
the upper‑left and upper‑right corners of the
monitor, and a thin line may appear along the
top. These blade edges are captured in the
view because of the wide‑angle camera lens
used in the video laryngoscope. This image acts
as a frame of reference during the intubation
process and ensures that the orientation of the
image is correct in the monitor.
RECOMMENdEd CHECKS
5. On the back of the monitor, remove the USB port cap, and then insert a USB flash drive into the port.
USB port
6. Ensure that the USB flash drive is detected by checking if the USB Flash Drive icon on the bottom of
the screen is displayed.