PROCEdURE 8. CLEAN THE SMART CABLE
Cleaning is critical to ensuring a component is ready for disinfection or sterilization. Failure
to properly clean the device could result in a contaminated instrument after completing the
disinfection or sterilization procedure.
When cleaning, ensure all foreign matter is removed from the surface of the device. This allows
the active ingredients of the chosen disinfection method to reach all the surfaces.
WARNING
Use this procedure in order to clean the Smart Cable. It is critical to remove all traces of contamination from
the component prior to completing disinfection or sterilization procedures. The Smart Cable is IPX7 compliant
and may be completely immersed in cleaning and disinfecting solutions. The use of protective caps is not
required during reprocessing.
To significantly reduce the amount of effort needed to clean the system, do not let contaminant(s) dry on
any system component. Bodily contaminants tend to become securely attached to solid surfaces when dried,
making removal more difficult.
Verathon has validated the products in this procedure for both chemical compatibility and biological efficacy
when cleaning the indicated component(s) as instructed in the Conditions column.
Table 9. Cleaning Methods for Smart Cables
PRODUCT LEVEL CYCLES* CONDITIONS
STERIS
®
Prolystica
®
2X
Concentrate
§
Cleaner 1500
Exposure: Prepare solution in warm water at ⅛ to ½ fl. oz.
per gallon (1–4 mL/L). Soak component for at least 3 minutes.
Before removing from solution, brush all surfaces using a
soft‑bristled brush, paying special attention to hard‑to‑reach
areas. Use a cotton swab for the camera window to avoid
damaging the window.
Rinse: Rinse for 3 minutes under warm running water. If
component is soaked for longer than 3 minutes, increase rinse
time in proportion to soak time.
Metrex
®
CaviCide
®
Cleaner 1500
Water temperature: 33–40°C (91–104°F)
Exposure: Spray all surfaces until drenched. Allow to remain
wet for 10minutes. Brush all surfaces. Rinse under running
water for 5minutes. Spray all surfaces until drenched. Allow to
remain wet for 10minutes.
Rinse: Rinse under running water for 5minutes. Fully immerse
in water and agitate for 3minutes. Before removing from
water, brush the cable with a soft‑bristled brush. Remove
from water and flush the connectors with running water
and a syringe. Fully immerse in fresh water and agitate for
3minutes. Rinse under running water for 2minutes.