28
USING THE DEVICE
Prior to using the device, set up the device according to the instructions in the previous chapter, and verify
the setup by completing the procedure Perform a Functional Check.
GlideScope systems are delivered nonsterile and require cleaning or disinfection prior to initial use.
WARNING
Before every use, ensure the instrument is operating correctly and has no sign of damage. Do
not use this product if the device appears damaged. Always ensure that alternative airway
management methods and equipment are readily available.
Report any suspected defects to Verathon
®
Customer Care. For contact information, visit
verathon.com/support.
WARNING
AVL video laryngoscopes are equipped with the Reveal™ anti‑fog feature, which reduces camera fogging
during the intubation procedure. To fully optimize the feature, you must allow the video laryngoscope to
warm up for 30‑120 seconds prior to use, depending on the ambient temperature and humidity of the
clinical environment. Full optimization of the anti‑fog feature is not necessary in order to use the device; if
desired, you may begin the intubation procedure immediately.
Note: If the video laryngoscope is stored in cold conditions, additional warming time may be required for
optimal performance of the anti‑fog feature.
Using the system consists of the following procedures:
• Procedure 1: Connect the Video Cable to the Monitor
• Procedure 2: Insert the Video Baton into the Stat (Single‑Use Only)
• Procedure 3: Prepare the GlideScope System
• Procedure 4: Intubate Using a Video Baton and Stat
• Procedure 5: Intubate Using the GlideScope Direct
• Procedure 6: Use the Record & Snapshot Features (Optional)
• Procedure 7: Use the Playback Feature (Optional)