45
Operations & Maintenance Manual: Cleaning & Disinfecting
10. Disinfect the video baton according to the conditions stated in Table 8. The exposure process and times
vary depending on the solution and the component.
If you are using pre‑moistened wipes, wipe down the component as often as needed to keep it
visibly wet throughout the specified exposure time. You may use multiple wipes if necessary.
IMPORTANT
11. If applicable, rinse the video baton according to the solution manufacturer’s instructions.
12. Dry the video baton as appropriate for the disinfectant you are using.
If you are using … Then …
A liquid disinfectant
Use a sterile cloth, hospital‑grade clean air, or a low‑temperature dryer
to dry the component.
Pre‑moistened wipes Allow the component to air dry completely.
13. Inspect the video baton according to the instructions in the following procedure, and then store the
disinfected video baton in a clean environment.
PROCEdURE 7. INSPECT THE VIdEO BATON
Before every use, ensure the instrument is operating correctly and has no sign of damage. Do
not use this product if the device appears damaged. Always ensure that alternative airway
management methods and equipment are readily available.
Report any suspected defects to Verathon
®
Customer Care. For contact information, visit
verathon.com/support.
WARNING
1. Visually inspect the video baton for signs of damage. Perform a routine inspection of the video baton
before and after every use to ensure that all endoscopic components are free of unintended rough
surfaces, sharp edges, protrusions, or cracks.