DIRECT INTUBATION TRAINER
The GlideScope Direct intubation trainer requires high‑level disinfection prior to use. For more information
about the risk classification of AVL system components, see Table 5 on page38.
Do not use metal or abrasive brushes, scrub pads, or rigid tools to clean the blade. The window that
protects the camera and light can be scratched, permanently damaging the device.
This component is heat‑sensitive, and exposing it to temperatures in excess of 60°C (140°F) will cause
damage to the electronics.
IMPORTANT
PROCEdURE 11. CLEAN & dISINFECT THE GLIdESCOPE dIRECT
Table 12. Cleaning & Disinfection Methods for the GlideScope Direct
PRODUCT
DISINFECTION
LEVEL
CONDITIONS
Enzymatic debridement
agent/detergent
Cleaner As per chemical manufacturer’s instructions
ASP
®
Cidex
®
OPA * High As per chemical manufacturer’s instructions
Metrex
®
MetriCide
®
Plus30 *
High As per chemical manufacturer’s instructions
Sultan
®
Healthcare
Sporox
®
II *
High As per chemical manufacturer’s instructions
STERIS
®
S40™
or S20™ *
Sterilization
Standard cycles in the following processors:
STERIS
®
SYSTEM 1
®
(outside U.S.)
SYSTEM 1E
®
(in U.S.)
SYSTEM 1 EXPRESS (outside U.S.)
SYSTEM 1 PLUS (outside U.S.)
ASP
®
Hydrogen
Peroxide Gas Plasma
Sterilization
STERRAD
®
100S (in U.S.),
STERRAD
®
100S short cycle (outside U.S.),
STERRAD
®
NX standard cycle, or
STERRAD
®
100NX standard cycle
* Tested for 500 compatibility cycles.
Tested for 100 compatibility cycles.
CLEAN THE GLIdESCOPE dIRECT
1. Ensure the video monitor has been turned off.
2. Detach the video cable from the monitor by turning the connector ring in the direction of the arrow.
Connector ring