Operations & Maintenance Manual: Cleaning & Disinfecting
PRODUCT LEVEL CYCLES* CONDITIONS
Metrex
®
EmPower
™
Cleaner 1500
Water temperature: 19–29°C (66–84°F)
Exposure: Prepare solution at 1ounce/gallon (7.8mL/L). Soak
component for 3minutes. Before removing from solution,
brush all surfaces and pay special attention to hard‑to‑reach
areas.
Rinse: Rinse for 3minutes under running water.
Tristel™ Trio
Wipes System
(outsideU.S.)
Cleaner 1500
Pre‑clean: Use 2 or more pre‑clean towelettes to remove all
visible contamination from the component.
Continue to the entry "Tristel™ Trio Wipes System" in Table 10
on page50.
Pro‑Line
Solutions
EcoZyme
®
Cleaner 1500
Water temperature: 19–29°C (66–84°F)
Exposure: Prepare solution at 1ounce/gallon (7.8mL/L) in
30–40°C (86–104°F) water. Soak component for 5minutes.
Before removing from solution, brush all surfaces and pay
special attention to hard‑to‑reach areas. Using a syringe, flush
the connectors.
Rinse: Rinse for 5minutes under running water. Using a
syringe, flush the connectors.
Metrex
®
CaviWipes™
Cleaner 1500
Water temperature: N/A
Exposure: Use towelette(s) to remove all visible contamination
from the component. Using fresh towelette(s), wet all surfaces and
allow to remain wet for 3minutes.
Rinse: N/A. Allow the component to thoroughly air dry.
Metrex
®
CaviWipes1™
Cleaner 1500
Use 3 or more towelettes to remove all visible contamination
from the component.
Continue to the entry "Metrex
®
CaviWipes1™" in Table 10 on
page50.
Wip’Anios
Premium
(outsideU.S.)
Cleaner 1500
Water temperature: N/A
Exposure: Use towelette(s) to remove all visible contamination
from the component. Using fresh towelette(s), wet all surfaces
and allow to remain wet for 5minutes.
Rinse: N/A. Allow the component to thoroughly air dry.
Clinell
®
Universal
Sanitizing
Wipes
(outsideU.S.)
Cleaner 1500
Water temperature: N/A
Exposure: Use towelette(s) to remove all visible contamination
from the component. Using fresh towelette(s), wet all surfaces
and allow to remain wet for 5minutes.
Rinse: N/A. Allow the component to thoroughly air dry.