Operations & Maintenance Manual: Cleaning & Disinfecting
PRODUCT
DISINFECTION
LEVEL
CYCLES* CONDITIONS
STERIS
®
S40™
orS20™
High 750
Standard cycles in the following processors:
STERIS
®
SYSTEM 1
®
(outside U.S.)
SYSTEM 1E
®
(in U.S.)
SYSTEM 1 EXPRESS (outside U.S.)
SYSTEM 1 PLUS (outside U.S.)
Tristel™ Trio
Wipes System
(outsideU.S.)
High 1500
Clean according to the entry "Tristel™ Trio Wipes
System" in Table 9 on page46.
Sporocidal: Apply two pumps of the activator foam
to a sporocidal towelette and manipulate towelette for
15seconds. Wet all surfaces of the component and allow
to remain wet for 30seconds.
Rinse: Use a rinse towelette to wipe all surfaces.
STERIS
®
Revital‑Ox™
Resert
®
XLHLD
†
Revital‑Ox™
Resert
®
HLD/
Chemosterilant
†
Resert
®
XL HLD
†
High 1500
Conditioning: 20°C (68°F) or higher
Water Temperature: 20°C (68°F) or higher
Exposure: Soak for 8minutes, ensuring that all air
bubbles are removed from the surface of the component.
Rinse: (1) 1‑minute immersion with agitation in
purewater. Ensure the connector is properly rinsed.
ASP
®
Cidex
®
OPA
High 1500
Conditioning: 20°C (68°F) or higher
Water Temperature: 20°C (68°F) or higher
Exposure: Soak for 10minutes, ensuring that all air
bubbles are removed from the surface of the cable.
Rinse: (3) 1‑minute immersions with agitation in purewater.
Metrex
®
MetriCide
®
OPAPlus
High 1500
Conditioning: 20°C (68°F) or higher
Water Temperature: 20°C (68°F) or higher
Exposure: Soak for 10minutes, ensuring that all air
bubbles are removed from the surface of the cable.
Rinse: (3) 1‑minute immersions with agitation in purewater.
Medivators
®
Rapicide
®
OPA/28
High 1500
Conditioning: 20°C (68°F) or higher
Water Temperature: 20°C (68°F) or higher
Exposure: Soak for 10minutes, ensuring that all air
bubbles are removed from the surface of the blade.
Rinse: (3)1‑minute immersions with agitation in pure water
Anios/Farmec
®
OPASTER
High 1500
Water Temperature: 20°C (68°F)
Exposure: Soak the component for 30 minutes in
OPASTER solution, ensuring that all air bubbles are
removed from the surface. Use the solution at full
strength.
Rinse: (3) 1‑minute immersions with agitation in pure
water. Ensure that any exposed connectors are properly
rinsed.