52
PRODUCT
DISINFECTION
LEVEL
CYCLES* CONDITIONS
Medivators
®
Rapicide
®
PA
30°C
High 100
Concentration: 850±100parts per million
Conditioning: 28–32°C (82–90°F)
Exposure: 5minutes in a Medivators
®
Advantage PlusAER
reprocessing system with the following configuration:
• Hookup: 2‑8‑002HAN Rev. B
• Parameter: 1‑24‑010CDISF
* Value indicates number of compatibility cycles tested on the component. Exceeding the recommended number of cycles may
affect the potential life of the product.
† This chemical may cause discoloration of metal components, but the discoloration does not affect system efficacy or
functionality.
‡ The minimum concentration shown in metric units is higher than the minimum concentration shown in U.S. units. This is
consistent with the manufacturer’s current instructions.
1. Ensure the Smart Cable has been properly cleaned, according to the procedure Clean the Smart Cable.
2. Prepare and condition the disinfection solution according to the solution manufacturer’s instructions and
the conditions stated in Table 10.
3. Disinfect the component according to the conditions stated in Table 10. The exposure process and times
vary depending on the solution.
If you are using pre‑moistened wipes, wipe down the component as often as needed to keep it
visibly wet throughout the specified exposure time. You may use multiple wipes if necessary.
IMPORTANT
4. Rinse the component according to the conditions stated in Table 10. The rinsing process and times vary
depending on the solution.
5. Dry the component as appropriate for the disinfectant you are using.
If you are using … Then …
A liquid disinfectant
Use a sterile cloth, hospital‑grade clean air, or a low‑temperature dryer
to dry the component.
Pre‑moistened wipes Allow the component to air dry completely.
6. Examine the component for any signs of damage. If damage is present, do not use the component, and
contact Verathon
®
Customer Care.
7. Store the component in a clean environment.