Troubleshooting 37
EN
Disconnection!
Medium priority
Hose system not con-
nected to unit prop-
erly or not connected
at all
Check connection of
hose to unit
Unit being operated
with open mask
(mask not on).
Put on mask or switch off
unit.
Device Error!
Pressure sensor
implausible
Medium priority
Drops of water in
pressure measuring
hose
Dry pressure measuring
hose as described in 6.2
Device Error!
Upper tempera-
ture limit
exceeded
Medium priority
Unit overheating due
to direct sunlight
Allow unit to cool down,
find a more suitable
place to set it up.
Medium priority
Unit operated out-
side permissible tem-
perature range
Display gone out
Acoustic signal for
at least
120 seconds, no
display visible.
Medium priority
No supply voltage
available.
Check connecting cord is
securely connected. Pos-
sibly check function of
supply socket by connect-
ing another piece of
equipment (e.g. a lamp).
The VENTIpower
battery has dis-
charged.
The VENTIpower must be
disconnected from the
VENTIlogic and
recharged. The
VENTIlogic can continue
to be operated via a sup-
ply socket.
Indicator Alarm Cause Rectification

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann VENTIlogic BiLevel ST and is the answer not in the manual?

Weinmann VENTIlogic BiLevel ST Specifications

General IconGeneral
BrandWeinmann
ModelVENTIlogic BiLevel ST
CategoryMedical Equipment
LanguageEnglish

Related product manuals