EasyManuals Logo

Weinmann VENTIlogic BiLevel ST User Manual

Weinmann VENTIlogic BiLevel ST
56 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #36 background imageLoading...
Page #36 background image
36 Troubleshooting
EN
Physiological alarms
Technical alarms
Indicator Alarm Cause Rectification
V
Tmin
Minimum respira-
tory volume set by
the doctor not
reached.
Low priority
Filter soiled. Clean/change filter
Mask leaking.
Adjust headgear/head-
band so that the mask
does not leak - replace if
necessary
Mask defective. Replace mask
Implausible settings.
Have settings checked
by your doctor.
IPAPmin (leakage)
Minimum therapy
pressure set by
doctor not
reached.
Low priority
Filter soiled. Clean/change filter
Mask leaking.
Adjust headgear/head-
band so that the mask
does not leak - replace if
necessary
Masal mask defective.
Replace mask
Implausible settings
Have settings checked
by your doctor.
Indicator Alarm Cause Rectification
Pressure measur-
ing hose discon-
nected!
Low priority
Pressure measuring
hose not properly fit-
ted or forgotten dur-
ing assembly
Check correct fitting/
presence of pressure
measuring hose
Device Error!
Excessive pressure!
Medium priority
Pressure sensor
defective
Have unit repaired
Pressure measur-
ing hose blocked.
Medium priority
Drops of water in
pressure measuring
hose
Dry pressure measuring
hose as described in 6.2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann VENTIlogic BiLevel ST and is the answer not in the manual?

Weinmann VENTIlogic BiLevel ST Specifications

General IconGeneral
BrandWeinmann
ModelVENTIlogic BiLevel ST
CategoryMedical Equipment
LanguageEnglish

Related product manuals