10 Troubleshooting
10.1
Error messages on the touchscreen
Error messages appear when there is a malfunc-
tion in the device. The following table provides
information about possible causes of an error mes-
sage and information about possible remedies.
Display/reaction Description of error Cause Remedy
“No signal” Measured values are not
transferred to the W
atch
Expert.
n W
atch positioned
incorrectly.
n Watch not wound.
n Microphone cable
not connected.
n Position the watch or
clockwork with its crown
against the signal sensor.
n Wind the watch.
n Connect the microphone
cable.
“Out of range” Measured values are not
transferred to the W
atch
Expert.
n Beat number is not
set correctly
.
n Set the beat number in
accordance with the watch
specifications.
n Set mode to “Automatic” .
10.2 Troubleshooting
Description of malfunction Cause Remedy
Implausible test result The watch is not positioned cor-
rectly
.
Reposition the watch (
Ä
page 45).
Test results are not being
printed.
Printer out of paper. Insert a new roll of paper. See the
manufacturer
’s documentation.
Unclean diagram Signal is too strong/weak. Adjust the signal strength using the
“Signal amplifier” touch button
(
Ä
page 49 ).
Troubleshooting
Troubleshooting