49
3062547_201802
31 CO
2
check/adjustmentbymatching
thepipelengths
CO
2
adjustmentby
matchingthepipelengths
HG 00 CO
2
adjustments
Factory setting: 3
1 2 3 4 5
Lower CO
2
valueHigher CO
2
value
WithcontractorparameterHG00(seechapter19"Contractorlevelparameters"),
theappliancescanbematchedtotheuesystembychangingthespeedfor
bothstagesandthereforetheCO
2
valuecanbechanged.
- Remove front and top casing.
- Remove the cap from the r.h. "ue gas" test port.
- Insert the test probe of the CO
2
measuring instrument into the "ue gas" test port
(approx. 70 mm).
- Turn the program selector switch on the BM module to "Emissions test" .
(Illuminated signal ring as status indicator ashes yellow.)
- Measure the CO
2
content and compare it against the values given in the stage 2 table
below.
- If required, alter the CO
2
content with parameter HG00.
- Modifying the parameter HG00 has also adjusted the speed for stage 1. For
checking purposes, the boiler must be set to stage 1.
- Switch HG28 burner operating mode from 2(two-stage) to 1(single stage).
- Measure the CO
2
content and compare it against the values given in the stage 1
table below.
Appliance open COB-15 COB-20 COB-29 COB-40
Set CO
2
value
Stage 2 12,7 ± 0,3%
Stage 1 12,7 ± 0,3%
O
2
set value
Stage 2 3,8 ± 0,4%
Stage 1 3,8 ± 0,4%
- If the value lies outside of the set value, the CO
2
-value must be set
via the pump pressure (see Commissioning without Wolf control accessories).
- Reset HG28 burner operating mode from 1 to 2.
CO
2
check/settinginthelower
load(stage1)
AdjustingtheCO
2
valueinthe
upperloadrange(stage2)
(emissionstestmode)